fda-pump FAIL

Analysis of the FDA benchmark centrifugal blood pump (Interlaboratory Study 2) from its supplied CAD (sources/pump_housing.step, sources/pump_rotor.step; authoritative, 1 CAD unit = 100 um, scaled by 0.1 to mm) and the transcribed packet sources/fda-pump-packet.md. The housing chamber model (the inner surface of the D60 x 9 chamber and its tubing: the curved D12 inlet tube and the D4.39 throat, di

gates

[A] parts_passn/a
[A] bom_completePASS
[A] joint_fidelityPASS
[A] clearancesPASS
[A] motion_clearPASS
[A] fluid_domainPASS
[B] export_agreementPASS
[B] performancen/a
[B] conservationPASS
[B] refinementFAIL
[C] source_hygienePASS
[C] boundaries_completePASS
[C] provenancePASS
[C] evidence_separationPASS

observables

idstatusvalue ± u_numericalmesh-resolvedcase
pressureRiseC1ok23665 ± 442.47 Pa-C1
pressureRiseC2ok48717 ± 694.44 Pa-C2
pressureRiseC3ok47785 ± 684.98 Pa-C3
pressureRiseC4ok7469.7 ± 723.51 Pa-C4
pressureRiseC5ok36240 ± 8078.1 PanoC5
pressureRiseC6ok35689 ± 559.16 Pa-C6
eulerPressureRiseC1ok19359 Pa-C1
eulerPressureRiseC2ok37943 Pa-C2
eulerPressureRiseC3ok37943 Pa-C3
eulerPressureRiseC4ok19359 Pa-C4
eulerPressureRiseC5ok37943 Pa-C5
eulerPressureRiseC6ok37943 Pa-C6

A value its mesh ladder does not resolve is not evidence of performance; report.md gives each prediction's validation basis.